Pensky-Martens Closed Cup Flash Point Tester
This Pensky-Martens closed cup flash point tester determines the flash point of petroleum products. It is suitable for determining the closed cup flash point of petroleum products with a flash point over 40℃.



Technical Specifications
Overview
Precision Flash Point Testing
This Pensky-Martens closed cup flash point tester is engineered to meet the stringent requirements of the GB/T261-2008 standard. Featuring a compact and durable stainless-steel design, it offers continuous, stepless heating control and precise temperature regulation for reliable laboratory results. The instrument is designed for ease of use, making it a dependable choice for metering and detecting units requiring accurate flash point determination of flammable liquids.
Technical Parameters
Electric Stirring Rates
- Power Supply
- AC (220±10%)V, 50Hz
- Total Power Consumption
- 650 W
- Heating Power Range
- 0 - 600 W (Continuously adjustable)
- Sample Heating Rate
- 1 - 12 °C/min (Controllable)
Oil Cup Specifications
Oil Cup Details
| Parameter | Value |
|---|---|
| Inner Diameter | 50.7 - 50.8 mm |
| Depth | 55.7 - 56.0 mm |
| Capacity | About 70 ml |
| Scribed Line Depth | 33.9 - 34.3 mm |
Operating Environment
- Environmental Limits
- Ambient Temp ≤ 35°CRelative Humidity ≤ 85%
Ignition and Measurement
Thermometer Scales
- -5°C to 110°C (0.5°C division)
- 20°C to 150°C (1°C division)
- 90°C to 370°C (2°C division)
- Ignition Source
- Gas / Civilian fuel (0.8mm bore diameter)